Coris > Products > Quatro VeT > FAQ
When the sample is not diluted enough, the migration and the washing step of the strip will not be optimal which could drive to two separate consequences, e.g. (1) a background noise at the test line, or (2) a general appearance of the strip which is more dark, more colorful, that could make the reading of specific signal more difficult. That’s the reason why it’s important to make a precise dilution of specimen as described in the notice of kit, because the specimen dilution has been properly assessed by Coris in order to guarantee the best characteristics results obtained (e.g. specificity and sensitivity) of the kit.
According to the "Instructions For Use", it is not allowed to exchange buffer between kits. If you have volume problems, please contact your distributor who is able to provide additional vials. Nevertheless, the buffer is the same within one product of different lot numbers (e.g. Rota-Strip buffers exchange between different lots will not give rise to any problems on the strips).
At present, 5 different kinds of buffers are used in Coris’products, i.e., Saline Buffer (for Influ-A Respi-Strip, Influ –A&B Respi-Strip), Extraction Buffer (for RSV-Strip), HC Dilution buffer (for Adeno Respi-Strip, Combi-Strip, Pylori-Strip, GastroVir-Strip), HydroK buffer (for dry swab Test restricted to Adeno Respi-Strip) and Sample Dilution Buffer for all the other products. Except the detection kits based on PCR method (T.cruzi OligoC-Test and Leishmania OligoC-Test) have their own running buffer.
Regarding the GastroVir Kit we received some particular comments claiming that the Control Line was too faint. The buffer has been improved to allow a better migration of the latex microspheres leading to a better and a more homogeneous signal. This improvement does not affect any characteristics of the kit (i.e. specificity and sensitivity).
The red arrow indicates the limit that could not be exceeded beyond during any incubation.
It could happen that in some cases operator exceeds the indicated limit line by using different but routinely used tubes in laboratories. However, it's still possible to security a validate test as long as the incubation level is still below 1cm (i.e. the green line indicated on the figure) while the gold conjugate is placed at 2cm from the bottom of the stick.
There are two different kinds of conjugates – a specific reagent for test line and a specific reagent for the control line - in the strip. A strong positive specimen has NO effect on the intensity of the control line onto the nitrocellulose of the strip. As long as the control line appears - with whatever intensity level - the test should be regarded as validated.
Control line is aimed to ensure that chromatography has been performed up to the top of the strip. It is not related with the result of the diagnostic itself.
Sometimes, the control line might be "slightly weak "without any precise reason. It could be only an effect of the sample migration and / or responsiveness at the control line. To sum up, the control line is aimed to validate the test using various specimen exists.
The product is stable during at least 15 weeks after first opening. 15 weeks is quite high: that means at least 2 tests can be performed per week. Stability depends on the storage conditions.
Immunochromatographic test uses nitrocellulose support which is sensitive to humidity.
That's why the desiccant is recommended to be kept in the tube and in the cap.
In each case, there is an internal control (upper line). This internal control will appear if the test is running well. However the test could be used after 15 or 20 weeks and the result can be valid if the control line appears.
The positive control should be used and diluted with the provided buffer of the respective kit. Then directly plunge the strip into the dilution. The mixture should be well homogenized before testing. Use the mixture if necessary and follow kit instructions.
It is various between positive controls forms, e.g. freeze-dried or liquid.
The dilution level of positive control determines the number of possible tests. Please reference it to the Instruction For Use of each positive control.
No. Maybe the sticker (with arrows) covering the bottom of the nitrocellulose is peeled off. In this case, the test is invalidated. Please perform the test with another strip.
If the sample does not migrate due to the presence of particles, centrifuge the diluted sample. Collect the culture supernatant and perform the test again.
Yes. It might cause the appearance of false positive signal on test.
The control line is aimed to ensure the user that the chromatography is performed up to the top of the strip. It does not give any indication on the diagnostic result. Whenever a signal is observed on the test line (weak or strong), it means that the sample is positive for the pathogen of interest.