Coris > Products > RESIST-1 MCR Multi > FAQ
According to the "Instructions For Use", it is not allowed to exchange buffer between kits. If you have volume problems, please contact your distributor who is able to provide additional vials. Nevertheless, the buffer is the same within one product of different lot numbers (e.g. Rota-Strip buffers exchange between different lots will not give rise to any problems on the strips).
At present, 9 different kinds of buffers are used in Coris’products, i.e., RE-A Buffer (for Influ –A&B Respi K-SeT), Extraction Buffer (for RSV-Strip), HC Dilution buffer (for Adeno Respi-Strip, Combi-Strip, Pylori-Strip, GastroVir-Strip, for Crypto-Giardia Duo-Strip, Giardia-Strip, and all cassette formats of the corresponding kits), HydroK buffer (for dry swab Test restricted to AdenoRespi K-SeT), ST-A buffer (for Clostridium K-SeT), LY-A Buffer (for RSV Respi K-SeT and for kits RESIST except RESIST CTX-M, RESIST-3 O.K.N. and RESIST-5 O.K.N.V.I.), LY-D Buffer (for RESIST-3 O.K.N. and RESIST-5 O.K.N.V.I.), LY-E Buffer (for RESIST CTX-M),BL-A Buffer (for HAT Sero K-SeT), and Sample Dilution Buffer for all the other products.
The test strip contains two different conjugate panels: one specific reagent for each test line and one specific reagent for the control line. A strongly positive sample has no effect on the intensity of the control line on the nitrocellulose strip. As long as the control line appears, regardless of its intensity, the test is considered valid.
Control line is aimed to ensure that chromatography has been performed up to the top of the strip. It is not related with the result of the diagnostic itself.
Sometimes, the control line might be "slightly weak "without any precise reason. It could be only an effect of the sample migration and / or responsiveness at the control line. The control line is aimed to validate the test using various specimen.
Once the pouch is opened, the test should be performed immediately. An unopened pouch may be kept at between 4 and 30°C and used until the shelf-life date indicated on the packaging.
The most plausible explanation for this invalid test (absence of a control line after execution) is a problem with the sample volume dispensed into the cassette well. The instructions for use specify a precise volume of liquid (3 drops, 100 µl, or 120 µl depending on the kit used) of the diluted sample. However, if you dispense less liquid or foam bubbles (dilution buffers containing detergent) of the diluted sample into the cassette, the conjugate may not be completely released from the drying zone, and the insufficient liquid volume will prevent proper and complete migration onto the cassette strip. We recommend that you perform the test again with the correct volume of liquid from the diluted sample preparation to restore proper migration. Please contact Coris (cc) if you encounter this technical problem again.
Yes. It might cause the appearance of false positive signal on test.
The control line is aimed to ensure the user that the chromatography is performed up to the top of the strip. It does not give any indication on the diagnostic result. Whenever a signal is observed on the test line (weak or strong), it means that the result of the test is positive for the pathogen of interest.
Yes. In the case of insufficient liquid volume, refer to the FAQ "The liquid does not migrate in the cassette, what should I do?" If too much liquid is deposited in the cassette well, the risk is that the conjugates in the strip will be diluted too much, which could reduce the test's detection capacity.