Innovative solutions
    for more effective diagnostics

Presentation

Products

Pathogens

Client Care

Science

IFU

Contact

Blog

NEWS

MEDICA Trade Fair 2017

November 13 to 16, 2017

 

Products Range

NEW-Resist Range

Oxa-48

KPC

RESIST-3 O.O.K.

RESIST-3 O.K.N.

Immunochromatographic Range

Pathogens detection

40/41 Adenovirus

Adenovirus

Clostridium difficile

Cryptosporidium & Giardia

Cryptosporidium

Escherichia coli O157

Rotavirus & 40/41 Adenovirus

Giardia

Helicobacter pylori

Influenza A & B viruses

Legionella

Respiratory Adenovirus

Rotavirus & Adenovirus

Rotavirus

Respiratory Syncytial Virus (RSV)

Streptococcus (group A)

Trypanosoma brucei gambiense (HAT)

Quatro-VeT (Veterinary)

Molecular Range

Leishmania

Pseudomonas aeruginosa

Scientific partners

PARADIS (Biowin)

CCH Fever (EC-FP7)

Chips 4 life (EC-FP7)

Nidiag (EC-FP7)

AquaVir (FP7)

Nareb (FP7)

Euroleish (H2020-MSCA-ITN)

ebolaMoDRAD (IMI-2)

 

Products  >  Human-Field  >  RESIST-3 O.K.N.  >  FAQ

RESIST-3 O.K.N. - FAQ

Why the kit doesn't contain loop in the sdandard packaging?

The kit doesn't contain bacteriological loops because that material is very commonly used in labs. In general labs can obtain competitive prices on the plastic disposable by buying them directly to their plastic disposables distributors.

Coris can supply kits containing loop upon specific request; however it's important to take into account of additional cost related to the plastic material, administration and shipment..

Is the buffer the same for all Coris BioConcept's K-SeT kits? If I need some more for my tests, may I use the one from another kit?

According to the "Instructions For Use" it is not allowed to exchange buffer between kits. If you face some volume problems, please contact your distributor. He will provide you with some more vials. Nevertheless, for one product, the buffer is always the same; so, for instance, using the HC Dilution Buffer from different Combi K-SeT lots will not give rise to any problems on the strips as long as those buffers are within their shelf-life range as indicated on the vial.

At present, 8 different kinds of buffers are used in Coris BioConcept’s products, i.e., RE-A Buffer (for Influ-A+B Respi K-SeT), Extraction Buffer (for RSV Respi K-SeT), HC Dilution buffer (for Adeno Respi K-SeT, Combi K-SeT, Pylori K-SeT), HydroK buffer (for dry swab Test restricted to Adeno Respi K-SeT), ST-A buffer (for Clostridium K-SeT), LY-A buffer (for OXA-48 K-SeT), BL-A Buffer (for HAT Sero K-SeT) and Sample Dilution Buffer (for Giardia K-SeT).

What happens when a clinical specimen is a strongly positive specimen? What about the control line

There are two different kinds of conjugates – a specific reagent for test line and a specific reagent for the control line - in the strip. A strong positive specimen has NO effect on the intensity of the control line onto the nitrocellulose of the strip. As long as the control line appears - with whatever intensity level - the test should be regarded as valid.

Why does the control line sometimes appear very weak while the test line is positive?

Control line is aimed to ensure that chromatography has been performed up to the top of the strip. It is not related with the result of the diagnostic itself.

Sometimes, the control line might be "slightly weak "without any precise reason. It could be only an effect of the sample migration and / or responsiveness at the control line. The control line is aimed to validate the test using various specimen.

How to use correctly K-SeT kits from Coris BioConcept?

  • Always store the reagents at the indicated storage temperature : between 4 and 30°C for all kits
  • Aways use gloves to manipulate kit components and samples
  • Always use the provided (Extraction, Dilution, …) Buffer even if the sample is already diluted (e.g. in a transport medium)
  • Always be sure that the solution sample – buffer is homogenous before testing
  • Always read the results when the dipstick into the device is wet (between 10 and 15 minutes depending on the test

How long can a pouch remain open before performing the test?

Once the pouch is opened, the test should be performed immediately. An unopened pouch may be kept at between 4 and 30°C and used until the shelf-life date indicated on the packaging.

The liquid does not migrate in the cassette, what should I do?

The most plausible hypothesis to explain the invalid test (no control line after running) is a problem of sample volume poured into the sample well of cassette.

In the notice, the instructions for use stipulates to dispense 3 drops of diluted sample delivered by the collection tube with dropper provided with the kit. However, if you put only 2 drops of diluted samples in tube, the conjugate may not be released completely from the pad where it is dried, and the reduced liquid volume does not allow a correct and complete migration on the strip in the cassette. We recommend to perform a new cassette test with 3 drops of preparation of diluted sample to recover the correct migration. Contact Coris (cc) if you encounter again this technical problem.

Is it a problem if the test has "run" more than the recommended time?

Yes. It might cause the appearance of false positive signal on test.

The signal of the test line is very weak in comparison with the control line. Do I have to consider that the result is positive?

The control line is aimed to ensure the user that the chromatography is performed up to the top of the strip. It does not give any indication on the diagnostic result. Whenever a signal is observed on the test line (weak or strong), it means that the result of the test is positive for the pathogen of interest.

Home   -   Site Map   -   Privacy Policy   -   Contact   -   FAQ                   All rights reserved CorisBioconcept SPRL 1996-2012©                           www.corisbio.com

Web traffic Optimisation
Universem

Design, Programming & hosting
Impact communication