According to the "Instructions For Use", it is not allowed to exchange buffer between kits. If you have volume problems, please contact your distributor who is able to provide additional vials. Nevertheless, the buffer is the same within one product of different lot numbers (e.g. Rota-Strip buffers exchange between different lots will not give rise to any problems on the strips).
At present, 8 different kinds of buffers are used in Coris’products, i.e., RE-A Buffer (for Influ –A&B Respi K-SeT), Extraction Buffer (for RSV-Strip), HC Dilution buffer (for Adeno Respi-Strip, Combi-Strip, Pylori-Strip, GastroVir-Strip, for Crypto-Giardia Duo-Strip, Giardia-Strip, and all cassette formats of the corresponding kits), HydroK buffer (for dry swab Test restricted to AdenoRespi K-SeT), ST-A buffer (for C.diff-Strip and Clostridium K-SeT), LY-A Buffer (for RSV Respi K-SeT and for kits RESIST), BL-A Buffer (for HAT Sero K-SeT), ST-1/ST-2 Buffers (for Strep-A Respi-Strip), and Sample Dilution Buffer for all the other products.
Regarding the GastroVir Kit we received some particular comments claiming that the Control Line was too faint. The buffer has been improved to allow a better migration of the latex microspheres leading to a better and a more homogeneous signal. This improvement does not affect any characteristics of the kit (i.e. specificity and sensitivity).
There are two different kinds of conjugates – a specific reagent for test line and a specific reagent for the control line - in the strip. A strong positive specimen has NO effect on the intensity of the control line onto the nitrocellulose of the strip. As long as the control line appears - with whatever intensity level - the test should be regarded as valid.
Control line is aimed to ensure that chromatography has been performed up to the top of the strip. It is not related with the result of the diagnostic itself.
Sometimes, the control line might be "slightly weak "without any precise reason. It could be only an effect of the sample migration and / or responsiveness at the control line. The control line is aimed to validate the test using various specimen.
Once the pouch is opened, the test should be performed immediately. An unopened pouch may be kept at between 4 and 30°C and used until the shelf-life date indicated on the packaging.
The most plausible hypothesis to explain the invalid test (no control line after running) is a problem of sample volume poured into the sample well of cassette.
In the notice, the instructions for use stipulates to dispense 3 drops of diluted sample delivered by the collection tube with dropper provided with the kit. However, if you put only 2 drops of diluted samples in tube, the conjugate may not be released completely from the pad where it is dried, and the reduced liquid volume does not allow a correct and complete migration on the strip in the cassette. We recommend to perform a new cassette test with 3 drops of preparation of diluted sample to recover the correct migration. Contact Coris (cc) if you encounter again this technical problem.
Yes. It might cause the appearance of false positive signal on test.
The control line is aimed to ensure the user that the chromatography is performed up to the top of the strip. It does not give any indication on the diagnostic result. Whenever a signal is observed on the test line (weak or strong), it means that the result of the test is positive for the pathogen of interest.
To date, the OXA-48 K-SeT kit has allowed to detect OXA-48 carbapenemase from 10 bacterial species (Citrobacter braakii, Citrobacter freundii, Enterobacter aerogenes, Enterobacter asburiae, Enterobacter cloacae, Enterobacter kobei, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Serratia marcescens). Moreover, 7 OXA-48 family variants have been detected (OXA-181, OXA-162, OXA-204, OXA-232, OXA-370 from K. pneumoniae, and OXA-204 and OXA-244 from E. coli.
The OXA-48 K-SeT test is designed to be used with a bacterial colony on solid medium. Other sample types are currently investigated.